This product cannot be shipped to your selected delivery address/country. Purchase has been disabled.
Indications
Total shoulder prosthesis for glenohumeral osteoarthritis (OA), revision arthroplasty with deficient rotator cuff, pseudoparalysis,
complex fracture (3 or more fragments) of humeral head in elderly patient.
Principes
• Lowering and medialisation of the joint’s centre of rotation
• Extension of adduction range of motion due to resurfacing of scapular pillar
• Option to convert fracture stem to reverse prosthesis without changing the stem
FEATURES
The DUOCENTRIC® product line integrates the biomechanical concepts of a reverse shoulder arthroplasty:
• The humeral stem (cemented or cementless) is available in 4 standard sizes and 2 revision sizes:
- requires 125° humeral cut relative to humeral shaft axis
- has a minimally invasive intramedullary shape with T-shaped cross-section and lateral fi ns. The lateral portion of the
cemented model has holes for tuberosity fi xation for complex fracture cases
• The humeral baseplate available in 2 sizes and 4 heights:
- accepts humeral insert
- has 5 possible settings for anteroposterior offset
- makes it possible to convert 140° cut with a fracture stem to a 125° cut required for reverse prosthesis (special model,
must be requested)
• Humeral insert available as standard model (3 thicknesses) or retaining model (2 thicknesses):
- secured to baseplate through lip and antirotation lug
- allows humerus height and lateralisation to be adjusted